SIGMA 2-16PK
No. 10166 (230 V/50;60 Hz),
No. 10165 (120 V/60 Hz),
No. 10164 (100 V/50;60 Hz)
- refrigerated version of SIGMA 2-16P
- Universal laboratory centrifuge; speed range up to 15.300 rpm
- Possibility of precooling the rotors during standstill; Refrigerant CFC-free (R 134 a)
SIGMA 2-16KC
No. 10170 (230 V/50 Hz),
No. 10169 (120 V/60 Hz),
No. 10168 (100 V/50;60 Hz)
- Model C with control unit Spincontrol Comfort
- Motorized lid lock
SIGMA 2-16KCH
No. 10167 (230 V/50 Hz)
- heated version of 2-16KC
- Temperature control +40°C up to +60°C depending on rotor and speed
Spincontrol Universal | Spincontrol Comfort | |
---|---|---|
Timer | 10 sec – 11 h 59 min | 10 sec – 11 h 59 min |
Display | LCD | Graphic LCD |
Plain text pop up windows | – | + |
Time increments [sec] | 10, 1 | 60, 1 |
Short run | + | + |
Continous run | + | + |
Speed increments [rpm] | 100, 1 | 10, 1 |
Permanent display preselected, actual values | + | + |
g-force increments [x g] | 10, 1 | 10, 1 |
Temperature increments [°C] | 1 | 1 |
Programs | 50 | 50 |
Program list shows all parameters | – | + |
Linear acceleration curves | 2 | 10 |
Quadratic acceleration curves | – | 10 |
Self made acceleration curves | – | 10 |
Linear deceleration curves | 2 | 10 |
Quadratic deceleration curves | – | 10 |
Brakeless | + | + |
Auto lid open, switchable | +, + | +, + |
Acoustic signal, switchable | +, + | +, + |
Optical display stillstand | + | + |
Signal length preselectable | – | + |
Languages | english | english, german |
Counting cycles centrifuge, each rotor | +, – | +, + |
Stillstand cooling, switchable | +, – | +, + |
All rotors pre programmed | + | + |
Max. radius reducible | – | + |
Calculation for liquid density >1,2 g/cm3 | – | + |
Start delay | – | + |
Control unit lockable | + | + |
4-digit code protection, free selectable | – | + |
Rotors & Accessories for the SIGMA 2-16PK, 2-16KC and 2-16KCH centrifuges
Order No. | Description | Max. rpm | Max. rcf |
---|---|---|---|
11124 | Swing-out rotor 24 x 1.5 / 2.2 ml reaction vials no. 15008, 15040, incl. hermetic lid | 14.000 1/min | 16.215 x g |
12148 | Angle rotor 24 x 1.5 / 2.2 ml reaction vials no. 15008, 15040, incl. hermetic lid | 15.300 1/min | 21.460 x g |
12132 | Angle rotor 30 x 1.5 / 2.2 ml reaction vials no. 15008, 15040, incl hermetic lid | 14.000 1/min | 21.913 x g |
12141 | Angle rotor 10 x 10 ml, for tubes no. 15000, 15010, 15019, incl. hermetic lid | 15.300 1/min | 19.890 x g |
12139 | Angle rotor 6 x 30 ml, for tubes no. 15029, 15030, 15032, 15024, incl. hermetic lid | 15.300 1/min | 20.414 x g |
12151 | Angle rotor 6 x 50 ml Falcon, incl hermetic lid | 9.000 1/min | 8.603 x g |
11190 | Swing-out rotor 4 x 100 ml complete, no. 11192, incl. buckets no. 13097, 4 glass tubes 100 ml no. 15100 and rubber cushions no. 16051 | 5.000 1/min | 3.773 x g |
11191 | Swing-out rotor 16 x 15 ml complete, no. 11192, incl. carriers no. 13012, 16 PS-tubes no. 15020 and rubber cushions no. 16015 | 5.000 1/min | 3.913 x g |
11192 | Swing-out rotor, without buckets | 5.000 1/min | 3.158/4.025 x g |
11122 | Swing-out rotor for 2 to 6 mikrotiterplates | 3.000 1/min | 1.238 x g |
11123 | Swing-out rotor for 2 to 6 mikrotiterplates | 4.000 1/min | 2.129 x g |
12071 | Angle rotor for 12 x 15 ml Falcon and 6 x 50 ml Falcon | 4.000 1/min | 2.254 x g |
12072 | Angle rotor for 20 x 15 ml, for tubes Ø 17 mm, length up to 120 mm | 4.000 1/min | 2.486 x g |
12073 | Angle rotor for 30 x 15 ml, for tubes Ø 17 mm, length up to 120 mm | 4.000 1/min | 2.486 x g |
11409 | Mikrohematokritplate for 24 capillaries 1.5 x 75 mm, 50 micro litre | 12.000 1/min | 14.489 x g |
12104 | Angle rotor for 12 PCR-stripes with 8 tubes 0.2 ml each, incl. hermetic lid | 14.000 1/min | 21.475 x g |
Order No. | Description | Suitable For: |
---|---|---|
13000 | Adapter for reaction vials 0.25/0.4 ml, no. 15014 | 12148, 12133, 11124 |
13002 | Adapter for reaction vials 0.5/0.75 ml, no. 15005 | 12148, 12133, 11124 |
13021 | Adapter for PCR-tube 0.2ml, no. 15042 | 12148, 12133, 11124 |
13009 | Multiple carrier for 5 RIA-tubes Ø 12 x 75 mm no. 15060 | 11192 |
13012 | Multiple carrier for 4 tubes 15 ml no. 15015, 15020, 15023, 15024 | 11192 |
13150 | Bucket with screw cap for 1 Falcontube 50 ml no. 15151 | 11192 |
13059 | Adapter for 1 tube 10 ml, no. 15000, 15010, 15039, max. Ø 16.2/19 x 75 – 85 mm | 12139 |
13060 | Adapter for 1 Falcontube 15 ml no. 15115 | 12151, 13150 |
13041 | Bucket with screw cap, max. length of tubes 110 mm, for round carriers no. 14029, 14030, 14031, 14032, 14033, 14034, 14035 | 11192 |
13152 | Multiple carrier for 2 culture tubes 15 ml | 11192 |
13079 | Adapter for 1 tube 50 ml | 12151 |
13097 | Bucket for 1 tube 100 ml no. 15100, 15102, 15103, 15106 | 11192 |
14029 | Adapter for 5 tubes 5 – 7 ml, max. Ø 12.5/15 x 65 – 105 mm | 13041 |
14030 | Adapter for 4 tubes 10 – 15 ml, max. Ø 16.2/17.5 x 80 – 110 mm | 13041 |
14031 | Adapter for 1 tube 25 – 30 ml, max. Ø 25/30 x 70 – 105 mm | 13041 |
14032 | Adapter for 1 tube 50 ml, max. Ø 35/38 x 70 – 105 mm | 13041 |
14033 | Adapter for 4 tubes 5 ml, max. Ø 13.5/17.5 x 70 – 110 mm | 13041 |
14034 | Adapter for 3 tubes 15 ml, max. Ø 17.3/19 x 80 – 110 mm | 13041 |
14035 | Adapter for 1 Sterilin-tube 30 ml with cap, flat bottom, Ø 25/31 x 65 – 95 mm | 13041 |
13037 | Multiple carrier for 6 x 5 – 15 ml 17/19 x 90 – 110, Monovettes | 11192 |
13064 | Adapter for 1 x 30 ml 25/31 x 65 mm Sterilin no. 3008 | 13150 |
2-16PK, 2-16KC, 2-16KCH | |
---|---|
Power supply [V/Hz] | 230 V/50 Hz 120 V/60 Hz (not KCH) 100 V/50;60 Hz (not KCH) |
Power consumption [W] | 720 |
Max. capacity [ml] | 4 x 120 |
Max speed [rpm] | 15.300 |
Min speed [rpm] | 100 |
Interference suppression | EN 61326 |
Dimensions: Height x Width x Depth | 310 x 550 x 570 |
Height with open lid | 705 |
Weight without rotor | 60 |
Kinetic energy max [Nm] | 9.962 |
Noise level at max speed (approx.) | |
Rotor 12148 | 65 dB (A) |
Rotor 11190 | 63 dB (A) |
Max. acceleration [sec] | |
Rotor 12148 | 27 |
Rotor 11190 | 22 |
Max. deceleration [sec] | |
Rotor 12148 | 18 |
Rotor 11190 | 17 |
Temperature setting range [°C] | -20°C – +40°C (KC/KCH) -10°C – +40°C (PK) |
Min rotor temperature at 23°C ambient temprature | < 4°C |